ADC Prosphyg 760

Littafin Amfani da ADC Prosphyg 760 Pocket Aneroid Sphygmomanometer

Model: Prosphyg 760

Brand: ADC

Gabatarwa

This manual provides comprehensive instructions for the proper use, care, and maintenance of your ADC Prosphyg 760 Pocket Aneroid Sphygmomanometer. Please read this manual thoroughly before using the device to ensure accurate readings and safe operation. This device is designed for professional and personal use in measuring blood pressure.

Bayanin Tsaro

Always follow basic safety precautions when using medical devices. Failure to do so may result in inaccurate readings or injury.

Abubuwan da aka gyara

Your ADC Prosphyg 760 Sphygmomanometer typically includes the following components:

Placeholder image for ADC Prosphyg 760 Sphygmomanometer components

Image showing the typical components of an aneroid sphygmomanometer, including the gauge, cuff, and inflation bulb. Please note: The provided image is a placeholder and does not depict the actual product.

Saita

Before taking a measurement, ensure the device is properly assembled and the patient is prepared.

  1. Cuff Selection: Choose the appropriate cuff size for the patient's arm circumference. The cuff should fit snugly but not too tightly.
  2. Matsayin Mara lafiya: Have the patient sit comfortably with their back supported and feet flat on the floor. The arm chosen for measurement should be supported at heart level.
  3. Wurin Cuff: Wrap the cuff around the bare upper arm, approximately 1 inch (2-3 cm) above the elbow crease. Ensure the artery marker on the cuff aligns with the brachial artery.
  4. Haɗa abubuwan haɗin gwiwa: Ensure the tubing securely connects the cuff to the aneroid gauge and inflation bulb.

Umarnin Aiki

Follow these steps to accurately measure blood pressure using your ADC Prosphyg 760.

  1. Close Air Release Valve: Turn the air release valve (usually a screw valve on the inflation bulb) clockwise until it is fully closed. Do not overtighten.
  2. Inflate Cuff: Rapidly inflate the cuff by squeezing the inflation bulb until the gauge reads approximately 20-30 mmHg above the patient's usual systolic pressure, or until the radial pulse is no longer palpable.
  3. Deflate Cuff Slowly: Slowly open the air release valve by turning it counter-clockwise. The needle on the gauge should drop at a rate of 2-3 mmHg per second.
  4. Listen for Korotkoff Sounds (if using stethoscope):
    • Matsi na Systolic: Note the reading on the gauge when you hear the first clear, tapping sound (Korotkoff Phase I).
    • Matsin Diastolic: Note the reading on the gauge when the sounds disappear completely (Korotkoff Phase V).
  5. Karatun Rikodi: Record both systolic and diastolic pressures.
  6. Cikakkun Deflate: Once readings are taken, fully open the air release valve to deflate the cuff completely.
  7. Wait Between Readings: If taking multiple readings, wait at least 1-2 minutes between measurements.

Lura: For accurate blood pressure measurement, it is recommended to use this device in conjunction with a stethoscope to auscultate Korotkoff sounds.

Kulawa da Kulawa

Proper care will extend the life and accuracy of your sphygmomanometer.

Shirya matsala

MatsalaDalili mai yiwuwaMagani
Maƙallin baya hura ko hura a hankali.Air leak in cuff, tubing, or bulb; valve not closed.Check all connections for tightness. Ensure air release valve is fully closed. Inspect cuff and tubing for cracks or holes.
Gauge needle does not return to zero.Gauge is out of calibration or damaged.Do not use. Contact ADC for calibration or repair.
Karatu marasa daidaituwa ko marasa daidaito.Improper cuff size or placement; patient movement; rapid deflation; gauge out of calibration.Ensure correct cuff size and proper placement. Instruct patient to remain still. Deflate cuff at 2-3 mmHg/sec. Check calibration.

Ƙayyadaddun bayanai

Garanti da Taimako

ADC products are manufactured to high standards and are backed by a limited warranty. For specific warranty details, please refer to the warranty card included with your product or visit the official ADC website.

For technical support, calibration services, or replacement parts, please contact ADC customer service:

When contacting support, please have your model number (Prosphyg 760) and purchase date available.

Takardu masu alaƙa - Prosphyg 760

Preview ADC Prosphyg Aneroid Sphygmomanometer: Umarnin don Amfani
Wannan takaddar tana ba da cikakkun umarni don ADC Prosphyg Aneroid Sphygmomanometer. Yana ba da cikakken bayani game da tsaftacewa da hanyoyin lalata, kulawar inganci, bayanin garanti, bin ƙa'idodi, jagororin zubarwa, bayanin na'urar, sabani, ma'anar alama, faɗakarwa gabaɗaya, aiki, hanyoyin aunawa, da kulawa da kiyayewa. Mahimmanci ga ƙwararrun kiwon lafiya da daidaikun mutane masu amfani da na'urar.
Preview Diagnostix Aneroid Sphygmomanometer Instructions for Use
Comprehensive instructions for using ADC Diagnostix Aneroid Sphygmomanometers, including pocket, palm, and clock types. Covers proper inflation, deflation, stethoscope placement, measurement techniques, care, maintenance, and warranty information.
Preview ADC Advantage Automatic Blood Pressure Monitor Instructions for Use
This manual provides comprehensive instructions for the ADC Advantage Automatic Blood Pressure Monitor models 6021N, 6022N, and 6023N. It covers setup, usage, measurement procedures, understanding blood pressure, error messages, and maintenance.
Preview ADC Advantage Connect Blood Pressure Monitor: Instructions for Use (Models 6024N & 6025N)
Comprehensive instructions for using the ADC Advantage Connect automatic home blood pressure monitors (Models 6024N and 6025N), including setup, measurement, app connectivity, and maintenance. Learn about features like MAM Average Mode, Irregular Heartbeat Detector, and Traffic Light Indicator.
Preview ADC Advantage Haɗa Umarnin Kula da Hawan Jini don Amfani
Cikakken jagora don amfani da ADC Advantage Haɗa na'urorin hawan jini (Model 6024N da 6025N), suna nuna haɗin ADC@Home don haɗin wayar hannu. Koyi game da saitin, aunawa, ayyukan ƙwaƙwalwar ajiya, gyara matsala, da garanti.
Preview ADC Advantage Connect 6024N/6025N Blood Pressure Monitor Instructions for Use
Comprehensive instructions for using the ADC Advantage Connect automatic home blood pressure monitors (Models 6024N and 6025N), including setup, measurement, app connectivity, and troubleshooting.